Chi tiết sản phẩm
Dải đo: điện áp không tiếp xúc 0 to ±3 kV DC hoặc peak AC.
Độ phân giải 1V
Độ chính xác: ±0.05% of full scale.
Tốc độ hồi đáp: ít hơn 50µs cho 1 kV step
Kích thước: 108 mm H x 223 mm W x 430 mm D
Trọng lượng: 5kg
Nguồn cung cấp: 180 đến 250 V AC
Cung cấp kèm theo: HDSD, dây nguồn
Chọn mua thêm đầu đo (Cần thiết chưa bao gồm theo máy)
Model 7000ER ( end-viewing )
Model 3800E-2 ( end-viewing)
Model 3800S-2 (side-viewing)
Chi tiết
Datasheet
MODEL 370 DC STABLE ELECTROSTATIC VOLTMETER
The Trek Model 370 is a precision electrostatic meter with the unique capability of making noncontacting surface voltage measurements in the range of 0 to ±3 kV DC or peak AC.
The 370 employs an electrostatic field-nulling technique which achieves high DC stability and high measurement accuracy even if the probe to measured surface spacing changes. This permits measurements of either stationary or moving surfaces without the need to establish fixed spacing to maintain accuracy.
An automatic gain feature of the Model 370 eliminates the need for manual adjustment when changing probes or when changing the probe to measured surface separation.
The Model 370 also features one-step, push-button zeroing. When the ZERO button on the front panel is pressed, the Model 370 automatically adjusts the output to zero volts when the probe is coupled to a known zero volt surface.
A precision voltage monitor provides a low-voltage representation of the measured electrostatic voltage for external monitoring purposes, or for use as a feedback signal in a closed-loop system.
An optional data acquisition module is available, featuring an IEEE-488 compatible interface, 14-bit resolution, 12-bit accuracy, and a programmable sampling period from 10 µs to 30 minutes.
The Model 370 can be operated on a bench top or, with optional hardware, in a standard 19-inch rack.
Show/Hide Product Technical Information
DATA ACQUISITION MODULE (OPTIONAL)
Provides data output using an IEEE-488 compatible interface.
Resolution
14 bit.
Accuracy
12 bit.
Sampling Period
Programmable from 10 µs to 30 minutes.
Storage
32 Kbytes (16 Kbytes of data words).
Acquisition Start Signal
External trigger or IEEE-488 Talk Enable or Group Trigger command.
Successive Data Acquisition
Internal timer or external trigger.
Interface Functions
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT0, C0, E2.
Address
Switch selectable between 0 and 30.
PERFORMANCE
All specifications are with a 3800E-2 probe with a probe-to-surface separation of 2 mm unless otherwise noted.
Measurement Range
0 to ±3 kV DC or peak AC
Measurement Accuracy
Voltage Monitor Output
Better than ±0.05% of full scale.
Voltage Display
Better than ±0.1% of full scale + 1 count, referred to the voltage monitor.
Speed of Response (10% to 90%)
Less than 50 µs for a 1 kV step.
Stability
Drift with Time
Less than 150 ppm/hour, noncumulative.
Drift with Temperature
Less than 100 ppm/°C.
PRODUCT FEATURES
Zero Control
A momentary push-button switch to produce zero volts output when the probe is coupled to a known zero volt surface.
Automatic Gain
The Model 370 automatically optimizes the AC response when changing the type of probe being used or when changing the probe-to-surface separation.
Digital Enable
An open collector, TTL compatible input to enable or disable the measurement. A TTL high (or open) will disable the measurement. A TTL low will enable the measurement.
Voltage Monitor Output
A buffered output providing a low-voltage replica of the measured voltage. 1/200, 1/300, 1/600, 1/1000 options available. (Offset Voltage and Output Noise specifications vary based on output scale factor.)
Scale
1/100th of the measured voltage.
Offset Voltage (at 1/100th scale)
Less than 10 mV.
Output Noise (at 1/100th scale)
Less than 20 mV rms (measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter).
Output Impedance
Less than 0.1 Ω.
Output Current Limit
±10 mA.
Voltage Display
4-character, 7-segment LED display.
Range
0 to ±3000 V.
Resolution
1 V.
Zero Offset
±1 count, referred to the voltage monitor.
Sampling Rate
2.5 readings per second.
GENERAL
Power Requirements
Line Voltage
Factory set for one of two ranges: 90 to 127 V AC or 180 to 250 V AC, at 50 to 60 Hz.
Power Consumption
100 VA, maximum.
Operating Conditions
Temperature
0°C to 40°C.
Relative Humidity
To 85%, noncondensing.
Altitude
To 2000 meters.
Probe-to-Surface Separation (recommended)
2 mm ±1 mm.
Dimensions
108 mm H x 223 mm W x 430 mm D
(4.25" H x 8.75" W x 17" D).
Weight
5 kg (11 lb).
Voltage Monitor Connector
BNC coaxial connector.
Digital Enable
BNC coaxial connector.
Ground Receptacle
Binding post.
AC Line Cord Receptacle
Standard IEC three-prong line cord receptacle with an integral fuse holder.
ACCESSORIES
Accessories Supplied
Operator’s manual, an IEC line cord for 90 to 127 V AC voltage source, or an unterminated power cord for a 180 to 250 AC line.
CERTIFICATION
TREK, INC. certifies that each Model 370 is tested and calibrated to specifications using measurement equipment traceable to the National Institute of Standards and Technology or traceable to consensus standards. A Certificate of Calibration accompanies each instrument when it is shipped from the factory.
All specifications are subject to change.
Trek ESD Brochure 2020